Gamma Calibrations (Cesium-137) and
Electronic Calibrations
A gamma field calibration (up to 1
R/hr) will be automatically performed on all instruments with exposure rate
scales unless otherwise specified. An electronic calibration will automatically
be performed on instruments with c.p.m. scales only. A two-point per range
calibration is performed. Instruments having only one common calibration
adjustment for all ranges will be adjusted on the most commonly used scale or
as indicated by the manufacturer.
As part of each calibration service,
and prior to all portable instrumentation calibrations, a battery integrity
check is made, and when needed, replacements are made at list price. Batteries
will be replaced when they fail to exceed 80% of their rating. Replacement,
brand and construction will be left to our discretion unless otherwise
requested.
Instruments having an "Integrate"
mode may have that feature calibrated or checked upon request only. The charge
for this service will be the same, per range, as the standard calibration for
that type of instrument.
Multiple detector instruments are
available at 75% of full calibration fee for each additional
deduction.
Instruments whose response exceeds
10% will be calibrated with a correction factor noted. Instruments whose
response exceeds 20% will be rejected from calibration and noted as such on the
certificate. An instrument sticker will not be issued in this instance,
however, the calibration fee will be charged.